Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

171

7.5. Manual Calibration Checks with Zero/Span Valves

Zero and span checks using the VARIOUS zero/span valve options available for the GFC 7000TA are
similar to that described in Section 7.3, except that the zero and calibration operations are initiated
directly and independently with dedicated buttons CALZ and CALS.

To perform a manual calibration check of an analyzer with a valve option installed, use the following
method.

STEP ONE

: Connect the sources of Zero Air and Span Gas as shown in Figures 7-3 through 7-6.

STEP TWO

: Perform the zero/span check.

Wait until STABIL

falls below 1.0

ppm.

This may take

several minutes.

The value of STABIL

may jump

significantly. Wait

until STABIL falls

below 1.0 ppm. This

may take several

minutes.

SPAN CAL M

STABIL=XXX.X PPM

CO2 =X.XXX

< TST TST >

ZERO

SPAN

CONC

EXIT

SAMPLE

STABIL=XXX.X PPM

CO2 =X.XXX

< TST TST > CAL

CALZ

CALS

SETUP

ZERO CAL M

STABIL=XXX.X PPM

CO2 =X.XXX

< TST TST >

ZERO

CONC

EXIT

SAMPLE

STABIL=XXX.X PPM

CO2 =X.XXX

< TST TST > CAL

CALZ

CALS

SETUP

EXIT

returns to the main

SAMPLE

display

ACTION:

Record the

CO2 readings

presented in the

upper right corner of

the display.

ACTION:

Record the

CO

2

readings

presented in the

upper right corner of

the display.

SAMPLE* RANGE = 500.000 PPM CO2

=X.XXX

< TST TST >

CAL

CALZ

CALS

SETUP

Scroll to the STABIL test

function.