Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

134

6.13.4.5. AIN Calibration

This is the sub-menu to conduct the analog input calibration. This calibration should only be necessary
after major repair such as a replacement of CPU, motherboard or power supplies. Activate the ANALOG
I/O CONFIGURATION MENU (See Section 6.13.1), then press:

Exit at any time to

return to the main

DIAG

menu

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

Continue pressing SET? until …

DIAG AIO

AIN CALIBRATED:

NO

< SET

SET>

CAL

EXIT

DIAG AIO

AIN CALIBRATED:

YES

< SET

SET> CAL

EXIT

Instrument

calibrates

automatically

Exit to return to the

ANALOG I/O

CONFIGURATION

MENU

STARTING FROM ANALOG I / O CONFIGURATION MENU

DIAG AIO

CALIBRATING A/D ZERO


CALIBRATING A/D SPAN

6.13.4.6. Configuring Analog Inputs (Option) Channels

To configure the analyzer’s external analog inputs option, define for each channel:

 gain (number of units represented by 1 volt)
 offset (volts)
 engineering units to be represented in volts (each press of the touchscreen button

scrolls the list of alphanumeric characters from A-Z and 0-9)

 whether to display the channel in the Test functions

To access and adjust settings for the external Analog Inputs option channels press: