3.7.4. Functional Check

1. After the analyzer’s components has warmed up for at least 30 minutes, verify that the software

properly supports any hardware options that were installed.

2. Check to make sure that the analyzer is functioning within allowable operating parameters.

Appendix C includes a list of test functions viewable from the analyzer’s front panel as well as
their expected values. These functions are also useful tools for diagnosing performance
problems with your analyzer (Section11.1.2). The enclosed Final Test and Validation Data sheet
(part number 04307) lists these values before the instrument left the factory.

To view the current values of these parameters press the following button sequence on the analyzer’s
front panel. Remember until the unit has completed its warm up these parameters may not have
stabilized.

RANGE

RANGE1

1

RANGE2

1

O

2

RANGE

2

STABIL

MEAS

REF

MR RATIO

PRES

SAMP FL

SAMP TEMP

BENCH TEMP
WHEEL TEMP

BOX TEMP

PHT DRIVE

SLOPE

OFFSET

TEST

TIME

SAMPLE

RANGE = 500.000 PPM

CO2 = XXX.X


< TST

TST >

CAL

SETUP

1

Only appears instrument is set

for DUAL or AUTO reporting
range modes.

2

Only appears if O

2

Sensor

Option is installed.

Refer to

Table 6-2 for

definitions of

these test

functions.

Toggle <TST TST> buttons

to scroll through list of

If your network is running a dynamic host configuration protocol (DHCP) software package, the

Ethernet feature will automatically configure its interface with your LAN. (See Section 6.11.6.2).
This configuration is useful for quickly getting an instrument up and running on a network.
However, for permanent Ethernet connections, a static IP address should be used. (See Section
6.11.6.3).