Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

247

Table 10-10: Analog Output Test Function - Nominal Values Voltage Outputs

FULL SCALE OUTPUT VOLTAGE

100MV 1V

5V

10V

STEP

%

NOMINAL OUTPUT VOLTAGE

1 0

0

0 0 0

2 20 20

mV 0.2 1 2

3 40 40

mV 0.4 2 4

4 60 60

mV 0.6 3 6

5 80 80

mV 0.8 4 8

6 100 100

mV 1.0 5 10

10.5.8.3. Analog Outputs: Current Loop

To verify that the analog outputs with the optional current mode output are working properly, connect a
250 ohm resistor across the outputs and use a voltmeter to measure the output as described in Section
6.9.4.2 and then perform an analog output step test as described in Section 6.9.3.

For each step the output should be within 1% of the nominal value listed in the table below.

Table 10-11: Analog Output Test Function - Nominal Values Current Outputs

OUTPUT RANGE

2 -20

4 -20

NOMINAL OUTPUT VALUES

STEP

%

CURRENT

V(250 OHMS)

CURRENT

V(250 OHMS)

1 0 2

mA

0.5V

4

1

2 20 5.6

1.4

7.2

1.8

3 40 9.2

2.3

10.4

2.6

4 60 12.8

3.2

13.6

3.4

5 80 16.4

4.1

16.8

4.2

6 100 20

5

20

5