Com port password security – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 169
Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
151
To initialize the modem press:
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG ALRM
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
COMMUNICATIONS MENU
ID INET
COM1
COM2
EXIT
EXIT
returns to
the previous
menu
SETUP X.X
COM1 MODEM INIT:AT Y
&D &H
<SET
SET>
EDIT
EXIT
SETUP X.X
COM1 INITIALIZE MODEM
<SET
SET> INIT
EXIT
SETUP X.X
COM1 MODE:0
SET>
EDIT
EXIT
SETUP X.X COM1 BAUD RATE:19200
<SET
SET>
EDIT
EXIT
EXIT
returns to the
Communications Menu.
SETUP X.X
INITIALIZING MODEM
<SET
SET> INIT
EXIT
6.15.2.7. COM Port Password Security
In order to provide security for remote access of the GFC 7000TA, a LOGON feature can be enabled to
require a password before the instrument will accept commands. This is done by turning on the
SECURITY MODE
(Mode 4, Section 6.11.5). Once the SECURITY MODE is enabled, the following items
apply.
A password is required before the port will respond or pass on commands.
If the port is inactive for one hour, it will automatically logoff, which can also be achieved with the
LOGOFF command.
Three unsuccessful attempts to log on with an incorrect password will cause subsequent logins to be
disabled for 1 hour, even if the correct password is used.
If not logged on, the only active command is the '?' request for the help screen.
The following messages will be returned at logon:
LOGON SUCCESSFUL - Correct password given
LOGON FAILED - Password not given or incorrect
LOGOFF SUCCESSFUL - Connection terminated successfully
To log on to the Model GFC 7000TA analyzer with SECURITY MODE feature enabled, type:
LOGON 940331
940331 is the default password. To change the default password, use the variable RS232_PASS issued
as follows:
V RS232_PASS=NNNNNN