Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

226

D1 – M/R Sensor Status

D2 – Segment Sensor Status

JP4 Connector to Opto-Pickup

Board

Figure 10-4:

Sync/Demod Board Status LED Locations

10.1.4.3. Relay Board Status LEDs

There are eight LEDs located on the Relay Board. The most important of which is D1 is the most
important LED, which indicates the health of the I

2

C bus. If D1 is blinking the other faults following LEDs

can be used in conjunction with DIAG menu signal I/O to identify hardware failures of the relays and
switches on the relay (See Section 6.9.2 and Appendix D).

Table 10-4: I2C Status LED Failure Indications

LED

FUNCTION

FAULT STATUS

INDICATED FAILURE(S)

D1

(Red)

I

2

C bus Health

(Watchdog

Circuit)

Continuously ON

or

Continuously OFF

Failed/Halted CPU
Faulty Motherboard, or Relay Board
Faulty Connectors/Wiring to/from Relay Board
Failed/Faulty +5 VDC Power Supply (PS1)

STATUS LED’s

DC VOLTAGE TEST

POINTS

RELAY PCA

Figure 10-5:

Relay Board Status LEDs

Advertising