Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

52

Figure 5-1:

Current Loop Option Installed on the Motherboard

5.2.1. Converting Current Loop Analog Outputs to Standard Voltage

Outputs

NOTE

Servicing or handling of circuit components requires electrostatic discharge

protection, i.e. ESD grounding straps, mats and containers. Failure to use ESD

protection when working with electronic assemblies will void the instrument

warranty.

See Section 12 for more information on preventing ESD damage.


To convert an output configured for current loop operation to the standard 0 to 5 VDC output operation:

1. Turn off power to the analyzer.

2. If a recording device was connected to the output being modified, disconnect it.

3. Remove the top cover