Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

96

6.8.7. Dilution Ratio

The dilution ratio is a software option that allows the user to compensate for any dilution of the sample
gas before it enters the sample inlet. Using the dilution ratio option is a 4-step process:

1. Select reporting range units: Follow the procedure in Section 6.8.6.

2. Select the range: Use the procedures in Sections 6.8.2 – 6.8.5. Make sure that the SPAN value

entered is the maximum expected concentration of the undiluted calibration gas and that the span
gas is either supplied through the same dilution inlet system as the sample gas or has an
appropriately lower actual concentration. For example, with a dilution set to 100, a 10 ppm gas
can be used to calibrate a 1000 ppm sample gas if the span gas is not routed through the dilution
system. On the other hand, if a 1000 ppm span gas is used, it needs to pass through the same
dilution steps as the sample gas.

3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluting gas and 1 part of

sample gas):


SAMPLE RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SETUP C.3

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP C.3

RANGE CONTROL MENU

MODE

SET UNIT

DIL

EXIT

DIL

only appears

if the dilution ratio

option has been

installed

SETUP C.3

DIL FACTOR: 1.0 GAIN

0 0 0 1 .0

ENTR

EXIT

Toggle these buttons to set the

dilution factor.

This is the number by which the

analyzer will multiply the CO

2

concentrations of the gas passing

through the reaction cell.

SETUP C.3

DIL FACTOR: 20.0 GAIN

0 0 2

0 .0

ENTR

EXIT

EXIT

ignores the

new setting.

ENTR

accepts the

new setting.

The analyzer multiplies the measured gas concentrations with this dilution factor and displays the result.

NOTE

Once the above settings have been entered, the instrument needs to be

recalibrated using one of the methods discussed in Section 7.