Accessing the diagnostic features, Signal i/o – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 139

Advertising
background image

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

121

6.13.1. Accessing the Diagnostic Features

To access the DIAG functions press the following buttons:



SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

ALRM

EXIT

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

DIAG

ANALOG OUTPUT

PREV

NEXT

ENTR

EXIT

SAMPLE* RANGE = 500.00 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS: 818

8 1 8

ENTR

EXIT

From this point

forward, EXIT returns

to the

SECONDARY

SETUP MENU

EXIT

returns

to the PRIMARY

SETUP MENU

DIAG

FLOW CALIBRATION

PREV

NEXT

ENTR

EXIT

DIAG

ELECTRICAL TEST

PREV

NEXT

ENTR

EXIT

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG

DARK CALIBRATION

PREV

NEXT

ENTR

EXIT

DIAG

PRESSURE CALIBRATION

PREV

NEXT

ENTR

EXIT

EXIT

returns

to the main

SAMPLE

display

DIAG

TEST CHAN OUTPUT

PREV

NEXT

ENTR

EXIT

6.13.2. Signal I/O

The signal I/O diagnostic mode allows reviewing and changing the digital and analog input/output
functions of the analyzer. See Appendix A-4 for a complete list of the parameters available for review
under this menu.

NOTE

Any changes of signal I/O settings will remain in effect only until the signal I/O

menu is exited. Exceptions are the ozone generator override and the flow sensor

calibration, which remain as entered when exiting.

Advertising