Theory of Operation

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

206

9.3.5.3. Zero/Span Valve Options

Any zero/span valve options installed in the analyzer are controlled by a set of electronic switches located
on the relay board. These switches, under CPU control, supply the +12VDC needed to activate each
valve’s solenoid.

9.3.5.4. IR Source

The Relay board supplies a constant 11.5VDC to the IR Source. Under normal operation the IR source is
always on.

9.3.5.5. Status LEDs

Eight LEDs are located on the analyzer’s relay board to show the current status on the various control
functions performed by the relay board (see Figure 10-14). They are:

Table 9-2: Relay Board Status LED’s

LED

COLOR

FUNCTION

STATUS WHEN LIT

STATUS WHEN UNLIT

D1 RED

Watchdog

Circuit

Cycles On/Off Every 3 Seconds under direct control of the
analyzer’s CPU.

D2

YELLOW

Wheel Heater

HEATING

NOT HEATING

D3

YELLOW

Bench Heater

HEATING

NOT HEATING

D4 YELLOW

Spare

N/A

N/A

D5 GREEN Sample/Cal Gas

Valve Option

Valve Open to CAL GAS
FLOW

Valve Open to SAMPLE GAS
FLOW

D6 GREEN Zero/Span Gas

Valve Option

Valve Open to SPAN GAS
FLOW

Valve Open to ZERO GAS FLOW

D7 GREEN Shutoff Valve

Option

Valve Open to CAL GAS
FLOW

Valve CLOSED to CAL GAS
FLOW

D8

GREEN

IR SOURCE

Source ON

Source OFF