Ml24xxa native commands gpib operation – Anritsu ML2430A User Manual

Page 105

Advertising
background image

same command as the one which changes to AUTO averaging, it will also
update the USER averaging number.

Example:

AVG A, AUTO, 64

This command will set the system to AUTO averaging and the USER av-
eraging number to 64. But, the Auto Averaging measurement system does
not use the USER averaging number.

NOTE

The AVG mnemonic can be sent to just change the
<mode> of averaging (MOV, REPEAT etc.) without sending
a number, but there must be a following COMMA to indi-
cate the <val> parameter is not being sent. See the first ex-
ample below.

The AVG mnemonic can also be sent to just change the
User Average Number <val> (1 to 512) without changing
the averaging mode, but there must be a COMMA to indi-
cate the <mode> parameter is not being sent. See the last
example below.

Examples:

AVG A, AUTO,

Change Sensor A to Auto Averaging (note following comma to indicate the
<val> parameter is not being sent).

AVG A, AUTO, 64

Change Sensor A to AUTO and the User Average Number to 64.

AVG A, MOV, 32

Change sensor A to MOVing average and the User Average number to
32.

AVG A, RPT,

Change sensor A to RePeaT average and keep the User Average number
as 32.

AVG A, , 128

Change sensor A User Average number to 128, but keep the previously
set averaging mode (note comma to indicate the <mode> parameter is
not being sent).

Query:

AVG? <S>

Returned

String:

AVG <S>,<MODE>,<VAL>

6-18

ML2430A OM

ML24XXA NATIVE COMMANDS

GPIB OPERATION