RL

Relative mode

Syntax:

RL<mode>

mode:

0
1
2

Remarks:

Relative mode permits any measurement result to be compared in dB or
percent to a reference value. Relative mode can be enabled using the cur-
rent power reading (RL1) or the previous reference level (RL2). Successive
measurements are displayed relative to this reference value. RL0 disables
relative mode.

RM

Range hold set

Syntax:

RM<val>EN

val:

0 to 5

Remarks:

Set the range to <val> and then sets range hold. A value of 0 selects Auto
Ranging, so that the range will change to take the best measurement auto-
matically.

Example:

To set the range to 3:

RM3EN

Related

Commands:

RH

RV

Service request mask value.

Syntax:

RV

Remarks:

Read service request mask value. The returned string format is: <integer
value>
Converting the integer value into an 8-bit binary number, each bit corre-
sponds to the Service Request mask bits.

SE

Select cal factor table

Syntax:

SE<table_number>EN

table

number:

0 to 9

ML2430A OM

6-113

GPIB OPERATION

HP 437B EMULATION COMMANDS