Remarks:

Triggers a single reading which is added to the internal digital filter and the
updated filter power level is returned on the GPIB. After a TR1 command,
the instrument returns to TR0 standby mode.

Related

Commands:

TR0, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2

TR2

Trigger with a settling delay

Syntax:

TR2

Remarks:

Triggers a new series of readings; enough to update the internal digital fil-
ter for a noise free reading at the current power level. The value is then re-
turned on the GPIB and returns to TR0 standby mode.

Related

Commands:

TR0, TR1, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2

TR3

Trigger free run

Syntax:

TR3

Remarks:

Sets the ML2430A Series back into free run mode on both channels.

Related

Commands:

TR0, TR1, TR2, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2

ZE

Zero sensors

Syntax:

ZE

Remarks:

Zero all connected sensors. The ML2430, when emulating the HP 438,
may take longer to Zero a sensor than the HP 438 itself. When performing
a zero the status byte should be used to identify when ZEROing is com-
plete.