Remarks:

Clears the cal factor table to one entry for 50 MHz at 100%, but does not
clear the identity of the table. The cleared table is automatically saved to
the sensor.

CFUID

Cal factor table identity update

SENSOR

Syntax:

CFUID <s>,<table number>,<identity>

s:

A or B

table

number:

1 to number of tables supported by the sensor type

identity:

Seven characters or until a message terminator will be read as the identity.

Remarks:

Updates the seven character identity string. This only affects the copy of

the cal factor table stored in the memory of the power meter. To take affect
and not be lost, the table must be saved to the sensor using the

CFUSAV

command.

Query:

CFUID? <s>,<table number>

Returned

String:

CFUID <s>,<table number>,<identity>

CFUNITS

Cal factor display units

SENSOR

Syntax:

CFUNITS <s>,<units>

s:

A or B

units:

% or PCT
dB or dBm

Remarks:

This command changes the display units of the calfactors between either
dB or pecentage. Note that this will also set the form the data is output
over the GPIB (or RS232) when requested.

Query:

CFUNITS? <s>

Returned

String:

CFUNITS <s>,<units>

CFULD

Cal factor table binary load

SENSOR

Syntax:

CFULD <s>,<table number>,<length>,<binary data>

s:

A or B

table

number:

1 to number of tables supported by the sensor type

length:

Length of message in bytes

6-24

ML2430A OM

ML24XXA NATIVE COMMANDS

GPIB OPERATION