If the source used does not provide a blanking output, the blanking signal may be
disabled as follows: select

System|Rear Panel|BNC. Select PORT until INPUT 1

(digital) is selected, then select

TTL Level to alter the active state of the blanking

signal expected. Setting the TTL Level to

LOW will allow the Source sweep to

progress without a Blanking signal. This is useful for third party sources or simple
VTO systems. If a Source Sweep is later selected which does provide a blanking
signal, remember to restore the polarity of this signal to

HIGH or an incorrect dis-

play will result. Operating a source sweep which has BANDSWITCH blanking de-
lays in it without an appropriate blanking signal may lead to glitches in the
resulting measurement at the bandswitch points (simple VTO systems do not
usually have bandswitch points).

Source sweep operation mode includes the following display controls:

CHANNEL
[GRMD]

Select Channel 1 or Channel 2. The channel selected will be dis-
played on the left of the screen, above the middle value of the y
axis.

DATA HOLD
[GRPIX]

This selects how the graph is displayed on the screen. Select from
NORMAL, Min&Max, Min, or Max.

With Min & Max selected, the minimum and maximum points for
each sample are shown. If Connect Points is ON (

SYSTEM|Graph-

ics|CONNECT), a vertical bar is drawn between these points. Min
displays only the minimum for this sample. Max displays only the
maximum for this sample.

MODE
[SRCMOD]

Source sweep mode: FREQUENCY or POWER.

START
[SRCSTFRQ
SRCSTPWR]

Sweep start frequency (MHz or GHz) or power (dBm)

4-24

ML2430A OM

SYSTEM MENU

OPERATION

NOTE

If either Min&Max, Min,
or Max is selected, the
display will keep the
“old” data and may ap-
pear stationary.

The

DATA HOLD mode in
use is displayed on the
left of the screen, below
the middle value of the
y axis.

ML2430A Series Power Meter

BNC Cables

Anritsu 68/69000 Synthesized Sweep Generator

DIGITAL INPUT 1

SEQ SYNC OUT

ANALOG INPUT 2

HORIZ OUT

DWELL

IN

10 MHz

REF OUT

EXT

ALC

IN

HORIZ

OUT

V/GHz

OUT

SEQ

SYNC

OUT

10 MHz

REF IN

SERIAL

I/O

AUX I/O

NOTE: USE SLOW BLO FUSES ONLY

!

!

CAUTION

> 18 kg

HEAVY WEIGHT

CAUTION

DO NOT OPERATE

WITH POWER CORD

UNGROUNDED

220V

2.5A

110V

5A

SELECT

LINE

INPUT

90 - 132VAC

180 - 264VAC

48 - 440Hz

IEEE-488 GPIB

110

220

Figure 4-5. Source Sweep Mode Interconnection Example