Gpib operation hp 436a emulation commands – Anritsu ML2430A User Manual

Page 184

Advertising
background image

I

Trigger without settling.

Remarks:

Triggers a single reading which is added to the internal digital filter and the
updated filter power level is returned on the GPIB. After an I command, the
instrument returns to standby mode (H).

OI

Identification

Remarks:

Ask for identification of current operating mode. Responds with “HP436.”

R

Free run mode

Remarks:

Sets the ML2430A Series back into free run mode to continuously take
measurements and output data.

T

Trigger with settling

Remarks:

Triggers a new series of readings; enough to update the internal digital fil-
ter for a noise free reading at the current power level. The value is then re-
turned on the GPIB and returns to standby mode (H).

V

Free run mode with settling

Remarks:

Sets the ML2430A Series back into free run mode to continuously take
measurements and output data after running a settling routine.

Z

Zero sensor

Remarks:

Zero out the noise from the sensor. When zeroing the ML2430 in HP 436
emulation mode, the 'Z1T' sequence followed by the '9+DI' described in
the HP 436 manual must be followed.

ML2430A OM

6-97

GPIB OPERATION

HP 436A EMULATION COMMANDS

Advertising