state:

ON or OFF

Remarks:

This will link the trigger set-up between Profile mode and Readout mode
so that the sample delay and the gate width will agree. A change to the
trigger set-up in either Readout or Profile system set-up will affect either
display mode.

Query:

LINK?

Returned

String:

LINK <state>

LLIM

Set Low limits

CHANNEL

Syntax:

LLIM <c>,<val>

c:

1 or 2

val:

Remarks:

Sets the low limit. The LLIMS command turns the limits on and off. It is not
necessary to enter the units as the limit value is checked against the dis-
played value.

Therefore, if the limits have been set for –10 dBm (LLIM 1, –10) and the
display units are subsequently changed from dBm to Watts, the system still
checks for the reading to rise above –10, even though the display units
type has been changed.

Related

Commands:

LLIMS

Query:

LLIM? <c>

Returned

String:

LLIM <c>,<val>

LLIMS

Turn on/off low limits

CHANNEL

Syntax:

LLIMS <c>,<state>

c:

1 or 2

state:

ON or OFF

6-52

ML2430A OM

ML24XXA NATIVE COMMANDS

GPIB OPERATION

Units

Min

Max

dBm

–99.99

+99.99

dBmV

–53.00

147.00

dB

µV

7.00

207.00

Watts

0.0

50.0

Volts

0.0

20.0

Advertising