Operation system menu – Anritsu ML2430A User Manual

Page 53

Advertising
background image

STOP
[SRCSPFRQ
SRCSPPWR]

Sweep stop frequency (MHz or GHz) or power (dBm)

NOTE

When the power meter is communicating with a signal
source/generator over the serial interface, if the source fre-
quency power level or the frequency itself is changed, the
source sweep display will be updated where appropriate.

Control

The Control menu adjusts cursor position and toggles the readout display in Pro-
file, Power vs. Time and Source Sweep modes, and provides control over display
scaling.

SWAP
<<
>>
[CUR]

SWAP selects which cursor to move. The presently selected cursor
is defined by a triangular marker at the top of the cursor line. Press
the << soft key to move the selected cursor left, and the >> soft key
to move the selected cursor right. Trigger delay and Gate Width are
related to the cursor positions. This feature aids in measurement of
pulsed signals. Changing either cursor, or adjusting the Gate width
value, updates both the cursors and the gate width value.

When enabled through the

System|Sound|CURSOR menu selec-

tion, if a cursor is moved into an illegal space such as the edge of
the screen or the end of valid data (trigger point on the left of the
screen) a warning beep will sound.

SCALE
[GRYT
GRYB]

Adjust the Y-axis of the graph using TOP dB Value and BOTTOM
dB Value parameters in the SCALE submenu. AUTO scale is based
on the min and max of the previous sweep.

READOUT
[GRDATA
GRDRQ]

The supplemental data readout is displayed or removed with the
READOUT soft key. The readout provides display data depending
on the graph mode and the data hold type selected, as shown
below.

Profile Mode

Data hold = NORMAL:

ML2430A OM

4-25

OPERATION

SYSTEM MENU

1 -10.61
2 -12.63

dB
dB
dB
mS
dB

-0.00

8.85

-17.41

AV

P
t

ACTIVE CURSOR

AVERAGE READING

BETWEEN CURSORS

Figure 4-6. Sample Readout Display, Profile Mode, Data Hold = NORMAL