Operation trigger menu – Anritsu ML2430A User Manual

Page 47

Advertising
background image

1. Connect to the rear panel digital input.

2. Select

Trigger|Setup|ARMING|Blanking ON.

3. Set the polarity of the blanking (System menu)

Example power meter reading: –9.16 dBm.

TYPE
[TRGTYP
GTTYP]

The Type selection (RISE or FALL) sets the trigger for a rising or
falling edge. When the trigger source is set to INTA or INTB (Inter-
nal A or B) the power meter triggers on a power level which is rising
or falling.

LEVEL
[TRGLVL
GTLVL]

The Level selection sets the internal trigger level. When the trigger
source is set to either INTA or INTB (internal sensor A or B) the
channel triggers on a power level (in dBm) given by the sensor.
This value must not take any cal factors or offsets that the meter
applies into account.

Trig 1

If Trigger Channel 1 SOURCE is set to Manual, this softkey initiates
a measurement for channel 1.

Trig 2

If Trigger Channel 2 SOURCE is set to Manual, this softkey initiates
a measurement for channel 2.

Trig 1&2

If Trigger Channels 1 and 2 SOURCE are set to Manual, this
softkey triggers both channels simultaneously.

ML2430A OM

4-19

OPERATION

TRIGGER MENU

NOTE

Effective range is to ap-
proximately –30 dBm
and is only active in DC
ranges 1 and 2.

-9

dBm

-50

TTL

GSM BURST

FROM GENERATOR

BURST TRIGGER SYNC

FROM GENERATOR

Figure 4-3. Typical Arming Diagram