Gpib operation ml24xxa native commands – Anritsu ML2430A User Manual

Page 140

Advertising
background image

Remarks:

The LLIMS command turns the limits on and off.

Related

Commands:

LLIM

Query:

LLIMS? <c>

Returned

String:

LLIMS <c>,<state>

MMRST

Min Max Tracking reset

CHANNEL

Syntax:

MMRST <c>

c:

1 or 2

Remarks:

This command resets the min/max values when in 'Readout' or 'Power vs..
Time' mode. In profile mode, this command is used to reset the channels
min/max values.

MNGDB

Output Min Graph Binary Data

DATA OUTPUT

Syntax:

MNGDB

Remarks:

Available in graph modes only. Outputs in binary form the min graph data
to the GPIB in the long integer form of 1024 bits per dB as a definite length
arbitrary block response data. The C programming example 'Binary output
decoding' on page 6-133 shows how to extract the binary data. The re-
sponse form is as follows :
MNGDB <#><length><number_of_bytes><data_byte_1><data_byte_2>
...<data_byte_n><\n>
<length> number of ASCII characters make up the number_of_bytes value
<number_of_bytes> number of bytes of data contained in rest of the string
<data_byte_n> four of these values makes up the long integer.
For example: FF FF D1 64 = –11932 As it is based on 1024 per dB, divide
by 1024 to get the dB value (–11.652).

MNGD

Output Min Graph Data

DATA OUTPUT

Syntax:

MNGD

Remarks:

Available in graph modes only. Outputs in ASCII form the min graph data.
The format is as follows:

ML2430A OM

6-53

GPIB OPERATION

ML24XXA NATIVE COMMANDS

Advertising