Ml24xxa native commands gpib operation – Anritsu ML2430A User Manual

Page 127

Advertising
background image

GPRST

Reset min/max graph

PROFILE SETUP

Syntax:

GPRST

Remarks:

When profile DATA HOLD mode is set to Min/Max points, this command is
used to reset the min/max values for each data point.

Related

Commands:

GRPIX

GRAUTO

Auto scaling

SYSTEM

Syntax:

GRAUTO

Remarks:

Auto scale for all graphic modes (Profile, Source Sweep and Power vs.
Time). This command auto scales the y axis only based on the currently
displayed data.

GRAVG

Average between profile cursors

PROFILE SETUP

Syntax:

GRAVG <state>

state:

ON or OFF

Remarks:

Turns on or off averaging between cursors. The data returned by the
GRDRQ command includes the average of all data points between the cur-
sors if GRAVG is turned ON.

Related

Commands:

GRDRQ

Query:

GRAVG?

Returned

String:

GRAVG <state>

GRCP

Connect points on profile

PROFILE SETUP

Syntax:

GRCP <state>

state:

ON or OFF

Remarks:

When set to ON, creates a line graph by connecting the profile data points
together. The default is ON.

Query:

GRCP?

6-40

ML2430A OM

ML24XXA NATIVE COMMANDS

GPIB OPERATION

Advertising