Gpib operation hp 437b emulation commands – Anritsu ML2430A User Manual

Page 202

Advertising
background image

Remarks:

Stores the present configuration to the selected register.

Example:

To store the current instrument configuration in register 2:

ST2EN

Related

Commands:

RC

SV

Save cal factor table

Syntax:

SV

Remarks:

Since the ML2430A stores the cal factor tables in the sensors, this
command forces the edits to a cal factor table to be saved to the sensor.
The operation can take a couple of seconds to complete.

TR0

Trigger hold mode

Syntax:

TR0

Remarks:

Sets both channels to trigger hold mode. It does not trigger until it receives
a TR1 or TR2 or GET (group executive trigger), *TRG or TR3 command. If
it receives a TR3 command it reverts back to the trigger mode it was in be-
fore the TR0 command was sent. If the REM line is low, this command has
no effect.

Related

Commands:

TR1, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2

TR1

Trigger immediate

Syntax:

TR1

Remarks:

Triggers a single reading which is added to the internal digital filter and the
updated filter power level is returned on the GPIB. After a TR1 command,
the instrument returns to TR0 standby mode.

Related

Commands:

TR0, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2

TR2

Trigger with a settling delay

Syntax:

TR2

ML2430A OM

6-115

GPIB OPERATION

HP 437B EMULATION COMMANDS