Ml24xxa native commands gpib operation – Anritsu ML2430A User Manual

Page 157

Advertising
background image

SENSTL

Set Sensor Settle Percentage

SENSOR

Syntax:

SENSTL <s>,<val>

s:

A or B

val:

0.01 to 10%

Remarks:

Sets how long the system waits for the signal to settle. The value parame-
ter is only used in DEFAULT measurement sensor mode. The settling time
allows some control over the tradeoff between speed, and the extent to
which a measurement has settled to its final value.

Query:

SENSTL? <s>

Returned

String:

SENSTL <s>,<mode>

SENTYP

Return sensor information

SENSOR

Syntax:

SENTYP <s>

s:

A or B

Remarks:

This command returns information on the selected sensor in string format:
<sensor type>,<sensor serial>

.

For example: Dual Diode, PBD16.

SRCMOD

Source Sweep Mode

SYSTEM

Syntax:

SRCMOD <mode>

mode:

FREQ
POWER

Remarks:

Determines whether the voltage sweep applied to the V/Ghz analog input
on the rear panel is interpreted as a frequency or power sweep. The x axis
of the graph on the display will be labeled appropriately.

Query:

SRCMOD?

Returned

String:

SRCMOD <mode>

SRCSPFRQ Source Sweep Stop Frequency

SYSTEM

Syntax:

SRCSPFRQ <freq_value>[units]

freq value:

10 kHz to 122 GHz

6-70

ML2430A OM

ML24XXA NATIVE COMMANDS

GPIB OPERATION

Advertising