Gpib operation hp 438a emulation commands – Anritsu ML2430A User Manual

Page 210

Advertising
background image

LN

Set linear units

Syntax:

LN

Remarks:

Changes the display to linear units (Watts).

OC

Set calibrator state

Syntax:

OC<state>

state:

0 (OFF)
1 (ON)

Remarks:

For example: OC0 (reference calibrator state set to OFF).

OI

HP Identity request

Syntax:

OI

Remarks:

The format of the returned string is:
<Company name>,<model>,<serial>,<firmware version>

OS

Set offset value

Syntax:

OS<val>EN

val:

–99.99 to +99.99 dB

Remarks:

Specifies the offset applied to the displayed value. Values can be entered
in 0.01 dB increments.

Example:

To set an offset of 10.13 to the displayed value:

OS10.13EN

PR

Preset the unit

Syntax:

PR

Remarks:

Presets the unit to the HP factory defaults. This command does not effect
the calibration factors stored in the sensor data tables. The defaults for the
HP 438A are:

Measurement mode = Sensor A

ML2430A OM

6-123

GPIB OPERATION

HP 438A EMULATION COMMANDS