Ml24xxa native commands gpib operation – Anritsu ML2430A User Manual

Page 153

Advertising
background image

and measurement channel setups.
When in Readout mode, the Channel 1 and Channel 2 values, and the
max/min values if present, are printed below the header.
In Profile and Power vs. Time modes, a graph is printed out below the
header with all the details shown.

PRNSEL

Select the type of printer

SYSTEM

Syntax:

PRNSEL <type>

type:

HP340
BJC80

Remarks:

Available printer selections are the HP DeskJet 340 and Canon BJC80.
Other 300, 500, 600 Series and later HP printers are typically compatible.
If the Canon BJC80 printer is selected, it must be set to EPSON LQ emu-
lation mode for proper operation. Refer to the printer manual for instruc-
tions on setting the emulation mode.

Query:

PRNSEL?

Returned

String:

PRNSEL <printer>

RCD

Range Calibrator data request

DATA OUTPUT

Syntax:

RCD <s>

s:

A or B

Remarks:

Returns the results from an ML2419A Range Calibrator run after the
Range Calibrator is disconnected from the power meter. While still
connected to the Range Calibrator, the results can be printed but not read
via GPIB. The results include values for each end of each sensor range
and the zero value, and are kept in non-volatile memory until the Range
Calibrator is connected and a calibration run again.

Response format:
'RCD<ws><sensor>,<state>[,<zero value>,<range 1 upper>,<range 1
lower>,........]'

<state>:
If state is FALSE, no data follows because there are no valid results for this
sensor available.

If state is TRUE, the results for the selected sensor are displayed in the fol-
lowing order:
zero value, range 1 upper, range 1 lower, range 2 upper, range 2 lower,

6-66

ML2430A OM

ML24XXA NATIVE COMMANDS

GPIB OPERATION

Advertising