Ml24xxa native commands gpib operation – Anritsu ML2430A User Manual

Page 133

Advertising
background image

val:

–150.0 to +250.0

Remarks:

It is not necessary to specify units as the displayed units are always as-
sumed. Profile and Source Sweep modes always use dBm, but Power vs.
Time mode can also use dB

mV or dBmV.

Query:

GRYT?

Returned

String:

GRYT <val>

GT0

Set to ignore the Group Execute Trigger (GET) GPIB com-
mon command

GPIB TRIGGER

Syntax:

GT0

Remarks:

The ML2430A Series will ignore the GET command or a *TRG.

Related

Commands:

*TRG, Group Execute Trigger (GET), GT1, GT2

GT1

Set 'GET' command to TR1 type (immediate) trigger

GPIB TRIGGER

Syntax:

GT1

Remarks:

When the ML2430A Series receives a GET or *TRG command, the system
will perform a TR1-type trigger command.

Related

Commands:

*TRG, Group Execute Trigger (GET), GT0, GT2

GT2

Set 'GET' command to TR2 type (settling delay) trigger

GPIB TRIGGER

Syntax:

GT2

Remarks:

When the ML2430A Series receives a GET or *TRG command, the system
will perform a TR2-type trigger command.

Related

Commands:

*TRG, Group Execute Trigger (GET), GT0, GT1

GTARM

Set profile trigger arming

TRIGGER

Syntax:

GTARM <state>

state:

ON or OFF

6-46

ML2430A OM

ML24XXA NATIVE COMMANDS

GPIB OPERATION