Hp 438a emulation commands gpib operation – Anritsu ML2430A User Manual

Page 211

Advertising
background image

Reference Oscillator = Off
Active entry channel = A
Measurement units = Watts
REL mode = off
Measurement parameters (set for Sensor A and Sensor B):

Cal Factor = 100.0%
Cal Adj = 100.0%
Offset = 0.00 dB

RA

Auto Range

Syntax:

RA

Remarks:

Sets the ML2430A Series to automatically select the correct range for the
measurement.

RC

Recall setup

Syntax:

RC<val>EN

val:

1 to 10

Remarks:

The ML2430A Series can store up to 10 instrument configurations for con-
venient recall. The configuration parameters stored are the same parame-
ters the ML2430A Series stores in its own *SAV and *RCL commands.
Therefore, RC is equivalent to *RCL, and ST is equivalent to *SAV.

Selecting Register 0 always restores the previous power meter configura-
tion, providing an expedient way to recover from an entry error.

RH

Range hold

Syntax:

RH

Remarks:

Hold the power meter in the current range. The differences in sensor
ranges must be taken into account when the ML2430 is being used to em-
ulate the HP 438.

RL

Relative mode

Syntax:

RL<mode>

6-124

ML2430A OM

HP 438A EMULATION COMMANDS

GPIB OPERATION