Hp 437b emulation commands gpib operation – Anritsu ML2430A User Manual

Page 195

Advertising
background image

Example:

To set the frequency of the input signal to 300 MHz:

FR300MZ

GT

Set group trigger

Syntax:

GT<mode>

mode:

0
1
2

Remarks:

0 = Ignore Group Execute Trigger (GET) command
1 = Trigger immediate response to ‘GET’ command
2 = Trigger with delayed response to ‘GET’ command
The GTn command configures what the device does when it receives the
‘GET’ command.
For example: GT1 sets the’ GET’ (Group Execute Trigger) to perform a TR1
type trigger.

Related

Commands:

TR

ID

Return identification string

Syntax:

ID

Remarks:

Returned format:
<company name>,<model>,<firmware version>

IDN?

HP 437B identity request

Syntax:

IDN?

Remarks:

Returned format:
<company name>,<model>,<firmware version>

KB

Calibration factor

Syntax:

KB<val><terminator>

val:

1.0 to 150.0%

6-108