APPENDIX A

Instrument Ac-
curacy:

< 0.5 %

Zero Set and
Drift:

< 0.5 % MA2420A; < 0.5 % MA2470A Series and
MA2440A Series. Percent of full scale in most sensitive
range, measured over one hour with maximum averag-
ing after one hour warm up at constant temperature.

Noise:

< 0.5 % of full scale in most sensitive range, measured
over a one minute interval with maximum averaging,
two standard deviations at constant temperature after
one hour warm up, typical. MA2470A Series, 20 pW
typical.

1.00 mW Power Reference

Frequency:

50 MHz nominal

Output Level:

1.00 mW, ± 1.2%/year, ± 0.9% RSS,
traceable to National Standards

VSWR:

< 1.04

Connector:

Type N female

SENSOR / CHANNEL CONTROL

Operating
modes:

Readout: dual channel. RF power or voltage.
Power vs. Time: single channel graphic of readout data
over adjustable time interval. RF power or voltage.
Profile: single channel RF peak power graphic display
for analysis of repetitive pulse or transient waveforms.
Source Sweep: Single channel graphic display syn-
chronized to an RF source.

Range Hold:

Current range or selectable 1 through 5. (6 with
Universal Power Sensor).

Averaging

Auto-averaging: Moving average increases averaging
at low power ranges.
Averaging Types: Auto, Manual (Moving, Repeat)
Manual Average Range: 1, 2, 4, ..., 512
Low-Level Averaging: Low, Medium, and High settings
apply post-average low pass filter to improve visibility
at high display resolution settings.

Limit Lines:

Fixed value high and low limits with audible, rear panel
TTL output, and/or visible Pass/Fail alarm indication.
Failure indication can be set to latch until cleared so
that a transient failure can be easily noticed.