Ml24xxa native commands gpib operation – Anritsu ML2430A User Manual

Page 137

Advertising
background image

val:

Remarks:

Sets the high limit. The HLIMS command turns the limits on and off. It is
not necessary to enter the units as the limit value is checked against the
displayed value. Therefore, if the limits have been set for –10 dBm (HLIM
1, –10) and the display units are subsequently changed from dBm to
Watts, the system will still check for the reading to rise above –10, even
though the display units type has been changed.

Example:

The high limit is set to –10dBm and turned ON. The display is in dBm. A
reading of –9.500dBm would pass. If the display is subsequently changed
to Watts, a reading of 112.2

mW would fail, because the DISPLAYED value

is higher than –10. Limit checking only uses the displayed value and does
not change its value even though the display units have changed.

Related

Commands:

HLIMS

Query:

HLIM? <c>

Returned

String:

HLIM <c>,<val>

HLIMS

Turn on/off High limits

CHANNEL

Syntax:

HLIMS <c>,<state>

c:

1 or 2

state:

ON or OFF

Remarks:

The HLIMS command turns the limits on and off.

Related

Commands:

HLIM

Query:

HLIMS? <c>

Returned

String:

HLIMS <c>,<state>

HOLD

Graph hold

CHANNEL

6-50

ML2430A OM

ML24XXA NATIVE COMMANDS

GPIB OPERATION

Units

Min

Max

dBm

–99.99

+99.99

dBmV

–53.00

147.00

dB

µV

7.00

207.00

Watts

0.0

50.0