e. Sensor OFFSETS are applied.
f. Relative is applied if it is set to on before switching to FAST mode, and if
display channel 1 is configured for a single sensor and dB units.
g. No other data output requests are processed while in FAST mode, ex-
cept for serial poll. FAST mode must be turned off, for example, to ask for
the identity data.
h. FAST mode will not operate when sent via RS232.

FBEEP

Fail Beep On/Off

SYSTEM

Syntax:

FBEEP <c>,<state>

c:

1 or 2

state:

ON or OFF

Remarks:

When ON, causes an audio beep every time the limits for the selected
channel fail. If FBEEP is ON, and FHOLD is ON, whenever the limits speci-
fied for the channel have been exceeded, a beep sounds once every sec-
ond until FHOLD is turned OFF, or the CLEAR key (CLR) is pressed. The
FAIL indication is not affected by the CLEAR key, and can only be cleared
by turning FHOLD off. If a limit fail happens again, the alarm sounds again.

Related

Commands:

FHOLD

Query:

FBEEP? <c>

Returned

String:

FBEEP <c>,<state>

FHOLD

Fail indicator Hold

CHANNEL

Syntax:

FHOLD <c>,<state>

c:

1 or 2

state:

ON or OFF

Remarks:

If the high or low limits fail, and this setting is turned on, the fail status con-
tinues until the command is turned off. All BNC outputs, beeps and dis-
plays continue to be in the 'fail' state until after the OFF is received.

Related

Commands:

FBEEP

Query:

FHOLD? <c>

Returned

String:

FHOLD <c>,<state>

6-38