Gpib operation hp 438a emulation commands – Anritsu ML2430A User Manual

Page 212

Advertising
background image

mode:

0
1
2

Remarks:

Relative mode permits any measurement result to be compared in dB or
percent to a reference value. Relative mode can be enabled using the cur-
rent power reading (RL1) or the previous reference level (RL2). Successive
measurements are displayed relative to this reference value. RL0 disables
relative mode.

RM

Range hold set

Syntax:

RM<val>EN

val:

0 to 5

Remarks:

Set the range to <val> and then sets range hold. A value of 0 selects Auto
Ranging, where the range will change to take the best measurement auto-
matically.

Example:

To set the range to 3:

RM3EN

Related

Commands:

RH

RV

Service request mask value

Syntax:

RV

Remarks:

Read service request mask value. The returned string format is: <integer
value>
Converting the integer value into an 8-bit binary number, each bit corre-
sponds to the Service Request mask bits.

SM

Status Message

Syntax:

SM

Remarks:

Returns the status message in the format:
AAaaBBCCccDDddEFGHIJKLMNOP<cr><lf>
where:
AA: measurement error code
aa: entry error code
BB: operating mode

ML2430A OM

6-125

GPIB OPERATION

HP 438A EMULATION COMMANDS