Gpib operation ml24xxa native commands – Anritsu ML2430A User Manual

Page 148

Advertising
background image

OFFFIX

Offset fixed value

SENSOR

Syntax:

OFFFIX <s>,<val>[units]

s:

A or B

val:

–99.999 to +99.999

units:

dB

Remarks:

The value added to the sensor if the offset type is set to FIXED.

Example:

To set the fixed offset for sensor A to -47 dBm:

OFFFIX A,-47DB

Query:

OFFFIX? <s>

Returned

String:

OFFFIX <s>,<val>

OFFTBL

Specify the table used to apply offsets to the sensor

SENSOR

Syntax:

OFFTBL <s>, <val>

s:

A or B

val:

1 to 5

Remarks:

If the Offset Type is set to TABLE, use this command to specify which of
the five offset tables to apply to the sensor.

The tables are a set of frequency-against-dB offsets. The offset value used
from the table depends on the setting of the frequency correction source. If
the source is FREQUENCY, the entered frequency is used to calculate the
offset from the table. If the frequency correction source is V/GHz, the fre-
quency value calculated from the supplied ramp input is used to calculate
the offset from the table.

If the frequency does not match any frequency in the table, interpolation is
used to calculate the correct offset.

NOTE

If the frequency is greater than the maximum frequency in
the table, the offset value from the maximum table fre-
quency is used. If the frequency is less than the minimum
frequency in the table, the offset from the minimum table
frequency is used. The frequency comparisons start from
the beginning of the table; if the entry is 0 Hz, this is
counted as the end of the table.

Query:

OFFTBL? <s>

ML2430A OM

6-61

GPIB OPERATION

ML24XXA NATIVE COMMANDS