GPIB PC CARD
SETUP

The following GPIB driver configuration set up is recommended for reliable GPIB
communication with the ML2430A Series power meter. The set up is expressed
in the terms used by the National Instruments GPIB ISA and PCI cards and driv-
ers for WIN95 and DOS.

GPIB Device Template

The ML2430A Series default primary address is 13. Separate device templates
for the primary address of each device can usually be set up separately. The set-
tings for the device template for the ML2430A Series are:

Terminate read on EOS

NO

Set EOI with EOS on write

YES

Type of compare on EOS

8 bit

EOS byte

0x0A (10 decimal)

Send EOI at end of write

YES

Readdressing

YES

Secondary address

NONE

GPIB Card Settings

The recommended GPIB card settings for use with the ML2430A Series are:

Terminate read on EOS

NO

Set EOI with EOS on writes

YES

Type of compare on EOS

8 bit

EOS byte

0x0A (10 decimal)

Send EOI at end of write

YES

System controller

YES

Assert REN when SC

YES

Enable Auto Serial polling

NO

NI card. Cable length for HS488

OFF

ML2430A OM

6-5

GPIB OPERATION

GPIB PC CARD SETUP