Gpib operation hp 436a emulation commands – Anritsu ML2430A User Manual

Page 182

Advertising
background image

6-13

HP 436A
EMULATION
COMMANDS

This section provides an alphabetical listing of the commands (mnemon-
ics) used to program the Model ML2430A Series Power Meter when in HP
436A Emulation mode. The emulation mode is set through the front panel
SYSTEM|Rear Panel|GPIB|MODE menu (see Chapter 4, Operation) or
through the GPIB command EMUL (page 6-96).

HP Emulation commands must not have a space between the command
header and the parameter, or commas between the parameters.

The format for HP Emulation commands is:

<command header><parameter 1><parameter

n>...

The end of the command text must be terminated with a line feed charac-
ter (0Ah, decimal 10) or a GPIB End of Transmission State (EOI), or both.

+

Disable cal factors

Enable cal factors

1, 2, 3,
4 & 5

Set sensor operating range

Remarks:

Range 5 is the highest power range, range 1 the lowest. (These are the
opposite to the ML2430A native mode ranges; that is, HP 436 range 5 sets
to ML2430A range 1, and HP 436 range 4 to ML2430A range 2, etc.)

When the ML2430A is being used to emulate the HP 436, the ranges are
reversed; that is, HP 436 range 1 (the lowest power range) is equivalent to
the ML2430A range 5, and HP 436 range 5 (the highest power range) is
equivalent to the ML2430A range 1. Refer to page 4-6 for more information
on sensor ranges.

9

Auto range

Remarks:

Sets the ML2430A Series to automatically select the correct range for the
measurement.

A

Watt

Remarks:

Set units to Watts. Turn relative mode off and do not allow relative.

B

ML2430A OM

6-95

GPIB OPERATION

HP 436A EMULATION COMMANDS