Gpib operation hp 437b emulation commands – Anritsu ML2430A User Manual

Page 198

Advertising
background image

state:

0 (OFF)
1 (ON)

Remarks:

For example: OF1 (Turn offsets ON).

OI

Return identification string

Syntax:

OI

Remarks:

Returned format:
<company name>,<model>,<firmware version>

OS

Set offset value

Syntax:

OS<val>EN

val:

–99.99 to +99.99 dB

Remarks:

Specifies the offset applied to the displayed value. Values can be entered
in 0.01 dB increments.

Example:

OS10.13EN

Set an offset of 10.13 to the displayed value.

PR

Preset the unit

Syntax:

PR

Remarks:

Presets the unit to the HP factory defaults. This command does not effect
the calibration factors stored in the sensor data tables.

RA

Auto Range

Syntax:

RA

Remarks:

Sets the ML2430A Series to automatically select the correct range for the
measurement.

RC

Recall setup

Syntax:

RC<val>EN

ML2430A OM