Gpib operation ml24xxa native commands – Anritsu ML2430A User Manual

Page 112

Advertising
background image

binary data:

Same data as that recieved by CFURD

Remarks:

Loads binary data into the cal factor table. This command will automatically
save the data to the sensor.

CFUPT

Preset cal factor table

SENSOR

Syntax:

CFUPT <s>,<table number>

s:

A or B

table

number:

1 to number of tables supported by the sensor type

Remarks:

Presets the cal factor table to the factory settings. The preset table is auto-
matically saved to the sensor. For a universal sensor, separate cal factor
tables exist for option 1 (when fitted) and normal operation. This function
will preset the table relating to the currently selected mode.

CFURD

Cal factor table binary read

SENSOR

Syntax:

CFURD <s>,<table number>

s:

A or B

table

number:

1 to number of tables supported by the sensor type
F for the factory default table

Remarks:

This command outputs the cal factor table in binary mode in the following
form:

CFURD<space><length of binary data>,<binary data>

<length of binary data>: Total length of the binary data message, in bytes,
after the comma.

<binary data>: Made up of :
a. 8 bytes; 7 for the identity, plus a NULL terminator
b. 2 bytes representing the number of table pair entries
c. The cal factor table data in binary form. The binary data is output in en-
tries which are frequency/factor pairs of six bytes. The frequency is held in
32768e-6LONG format and the cal factor in 1024INT format.

To convert these into real numbers the first four bytes of an entry are read
into a LONG variable, cast to a float and then divided by 32768e-6 to give
a frequency. The last two bytes are then read into the low bytes of a
LONG then cast to a float and divided by 1024. The C programming ex-
ample 'Binary output decoding' on page 6-133 shows how to extract the
binary data.

ML2430A OM

6-25

GPIB OPERATION

ML24XXA NATIVE COMMANDS