Gpib operation ml24xxa native commands – Anritsu ML2430A User Manual

Page 154

Advertising
background image

range 3 upper, range 3 lower, range 4 upper, range 4 lower, range 5 up-
per, range 5 lower.

REL

Relative control

CHANNEL

Syntax:

REL <c>,<mode>

c:

1 or 2

mode:

0 Turn OFF
1 Turn ON and reference
2 Turn ON, use old references if not first time.

Remarks:

Turns relative ON or OFF, or references the zero point.
REL1 and REL2 toggle between relative and absolute measurements.
Sending the RELx command when in dB mode will make the meter apply
the last used RELATIVE value. This relative value is used thereafter until it
is replaced by another one in the same manner. This allows the user to re-
fer to a previously referenced value, without the meter resetting itself back
to a 0.00 display.

Query:

REL? <c>

Returned

String:

REL <c>,<mode>

RFCAL

Turn RF reference calibrator ON or OFF

CALIBRATION

Syntax:

RFCAL <state>

state:

ON or OFF

Remarks:

Turns on or off the RF reference calibrator.

Query:

RFCAL?

Returned

String:

RFCAL <state>

RGH

Range Hold Sensor

SENSOR

Syntax:

RGH <s>[,<val>]

s:

A or B

val:

0 to 5 (0 to 6 for universal power sensor only)
(0 = AUTO)

ML2430A OM

6-67

GPIB OPERATION

ML24XXA NATIVE COMMANDS

Advertising