Gpib operation hp 437b emulation commands – Anritsu ML2430A User Manual

Page 196

Advertising
background image

terminator:

%
PCT
EN

Remarks:

The calibration factor compensates for mismatch losses and effective effi-
ciency over the frequency range of the power sensor.

Example:

KB99.9%

KB99.9PCT

KB99.9EN

LG

Set log units

Syntax:

LG

Remarks:

Changes the display to log units (dBm).

LH

Set high limit

Syntax:

LH<val>EN

val:

–99.999 to +99.999 (dBm only)

Remarks:

Sets the high limit.

Example:

LH30.00EN

LL

Set low limit

Syntax:

LL<val>EN

val:

–99.999 to +99.999 (dBm only)

Remarks:

Sets the low limit.

Example:

LL20.00EN

LM

limits check state

Syntax:

LM<state>

state:

0 (off) or 1 (on)

Remarks:

Turns limit checking on or off.

ML2430A OM

6-109

GPIB OPERATION

HP 437B EMULATION COMMANDS