Ml24xxa native commands gpib operation – Anritsu ML2430A User Manual

Page 107

Advertising
background image

Syntax:

BAUTS <state>

state:

ENABLE or DISABLE

Remarks:

Enable/disable the battery auto power shut off.
NOTE: Although GPIB is not available under battery operation, the state of
this parameter can be changed for later use.

Query:

BAUTS?

Returned

String:

BAUTS <state>

BAUTT

Battery Auto shut off after

x minutes

SYSTEM

Syntax:

BAUTT <val>

val:

10 to 240 minutes

Remarks:

Automatically turns the unit off after

x minutes when operating on battery

power.
NOTE: Although GPIB is not available under battery operation, the state of
this parameter can be changed for later use.

Query:

BAUTT?

Returned

String:

BAUT <val>

BUFF

GPIB response buffering enabled

SYSTEM

Syntax:

BUFF <s>

s:

ON
OFF

Remarks:

If BUFF is ON: In the ML243X native mode, 488.2 GPIB operation, when a
request for data is made the response is put in an output buffer ready to be
read by the controller. If another data request is made and the previous
data has not been read out of the output buffer; the new data is queued af-
ter the original request. In this mode of operation the GPIB response buff-
ering enable is ON, and following the 488.2 specifications when ever a
request for data is made the response should be read.

If BUFF is OFF: In this mode when ever a request for data is made, (ex-
cept by serial poll) the output buffer is cleared and the only data in the
output queue will be the response to the last data request made. The out-
put buffer is cleared once a valid GPIB data request command has been
recognised.

6-20

ML2430A OM

ML24XXA NATIVE COMMANDS

GPIB OPERATION

Advertising