<11. MAINTENANCE>

11-2

IM 01F06A00-01EN

11.1 Changing the Converter and the Terminal Box Orientation

The converter and the terminal box can be changed in four directions with respect to the fl ow direction.

Integral Type Vortex Flowmeter

Remote Type Vortex Detector

Amplifier Unit

Indicator

Converter Cover

*

*

Bracket
Mounting
Bolt (two)

Body

Bracket

Vortex Shedder
Assembly

Hexagon
Mounting
Bolt (four)

Wire

Terminal Box Cover

Terminal Box

Bracket

Hexagon Mounting

Bolt (four)

Bracket Mounting
Bolt (two)

Wire

Vortex Shedder
Assembly

Body

Wire Color Terminal

Red

White

A
B

Wire Color Terminal

Red

White

A
B

Locking
Screw
(WAF: 1.5mm)

Shielded Cover

Locking
Screw
(WAF: 1.5mm)

Locking Screw


For the explosion
protected type
Flameproof (TIIS,
ATEX, IECEx)
Intrinsically safe
(SAA)

Locking Screw


For the explosion
protected type
Flameproof (TIIS,
ATEX, IECEx)
Intrinsically safe
(SAA)

WAF:

3mm

WAF:

3mm

F1101.ai

<1> Remove the converter cover. In case of the explosion

proof type cover removal, loosen the Locking Screw
(WAF: 3mm).

<2> For indicator and amplifier unit removal, refer to Section

11.2 “Indicator Removal and Rotation” and Section 11.3
“Amplifier Unit Removal”.

<3> Disconnect the vortex shedder assembly lead-wires from

the converter. In case of the explosion proof type, loosen
the Locking Screw (WAF: 1.5mm).

<4> Remove the bracket mounting bolts and remove the

converter and bracket from the flowmeter body. The
bracket applies to the 1 (25mm) to 4 (100mm) inch
flowmeters.

<5> Remove the hexagon mounting bolts in case of 90-degree

turn.

<6> Turn the converter to the desired orientation. When

reassembling the converter, reverse the above procedure.

<7> After changing the direction, make sure the impedance

between the earth terminal and the metal part of body,
vortex shedder assembly or bracket is 100ŸRUOHVV

<1> Remove the terminal box cover. In case of the explosion

proof type cover removal, loosen the Locking Screw
(WAF: 3mm).

<2> Disconnect the vortex shedder assembly lead-wires from

the terminal box. In case of the explosion proof type,
loosen the Locking Screw (WAF: 1.5mm).

<3> Remove the bracket mounting bolts and remove the

terminal box and bracket from the flowmeter body.

The bracket applies to the 1 (25mm) to 4 (100mm) inch
flowmeters.

<4> Remove the hexagon mounting bolts in case of 90-degree

turn.

<5> Turn the terminal box to the desired orientation. When

reassembling the terminal box, reverse the above
procedure.

<6> After changing the direction, make sure the impedance

between the earth terminal and the metal part of body,
vortex shedder assembly or bracket is 100ŸRUOHVV