<14. EXPLOSION PROTECTED TYPE INSTRUMENT>

14-13

IM 01F06A00-01EN

(5) Installation Diagram

Intrinsically Safe

F1410.ai

+

-

+

+

-

+

+

-

+

-

+

-

+

-

+

-

+

-

+

-

+

-

[Remote type]

DYA (converter)

SUPPLY

PULSE

A

B

T (*1)

C

A

B

T

DY (flowmeter)

SUPPLY

PULSE

Safety barriers

Safety barriers

[Integral type]

DY-N (flowmeter)

Hazardous Location
(Zone 0)

Non Hazardous
Location

Non Hazardous
Location

DYC: Signal cable

(*1) Wire for T terminal
With temperature sensor type : installed
Without temperature sensor type: not installed

Electrical parameters:
Ui=30V

Ii=165mA

Pi=0.9W

Ci=37nF

Li=0mH

Installation Requirements between Flowmeter, Converter and
Safety Barrieres:
Uo ” Ui Io ” Ii Po ” Pi Co • Ci1Ccable

Lo • Li1Lcable
Uo, Io, Co, and Lo are parameters of safety barrier.
Ccable and Lcable are the maximum allowable cable
parameters.

Hazardous Location
(Zone 0)

WARNING

• The safety barrier must be SAA certifi ed.
• Input voltage of the safety barrier must be

less than 250Vrms/Vdc.

• In any safety barriers used the output current

must be limited by a resistor ‘R’ such that
Io=Uo/R.

Type n

+

-

+

-

+

-

+

-

+

-

+

-

+

-

+

-

[Remote type]

A

B

T (*1)

C

A

B

T

DY (flowmeter)

[Integral type]

Non Hazardous
Location

Non Hazardous
Location

DYC: Signal cable

(*1) Wire for T terminal
With temperature sensor type : installed
Without temperature sensor type: not installed

SUPPLY

PULSE

SUPPLY

PULSE

Power

Supply

Power

Supply

Receiver

Receiver

DYA (converter)

DY-N (flowmeter)

General Purpose

Equipment

General Purpose

Equipment

Hazardous Location
(Zone 2)

Hazardous Location
(Zone 2)

F1411.ai

+

-

+

+

-

+

WARNING

Output voltage of power supply must not exceed
30Vdc and will void the certifi cation.

(6) Name Plate

Intrinsically safe and Type n

WARNING

• The

specifi cations of Intrinsically safe

type and Type n are described (engraved)
on these name plates. For use of SAA,
select a specifi cation either Intrinsically
safe type (Ex ia) or Type n (Ex n), then
erase unnecessary specifi cation at the
customer side.

Once a specifi cation has been selected,

never it can be changed again. (ex. Never
use the fl owmeter as Type n if it has selected
Intrinsically safe type fi rstly.)