<6. PARAMETERS>

6-14

IM 01F06A00-01EN

[C40:TIME UNIT] TIME UNIT

Select time unit from “/s(0), /m(1), /h(2), /d(3)”

[C45:FLOW SPAN] Flowrate span

Set the required span with a numerical value.

[C50:DAMPING] Damping time constant

Set damping time constant values from 0 to 99s.

(4) Item D (AUX. SETUP)

These items are for Auxiliary setup.
A value in “( )” is the data corresponding to indicator.

[D10:LOW CUT] Low-cut fl owrate

NOTE

For D10 setting, be sure to set “NOMINAL SIZE”
in E10 fi rstly.

Set to noise elimination or zero fl ow in the low
fl owrate (or low frequency) range. The settable
range for low cut fl owrate is more than half-
minimum fl owrate.

NOTE

In case that Reynolds adjustment (H25), Gas
expansion correction (H30) or select a break
point correction (H40) is necessary to set,
D10: LOW CUT must be set after the items
for compensations (H25, H30, H40) are set to
“ACTIVE”.

[D20:TEMP UNIT] Fluid temperature unit at
operating conditions

Select temperature unit at operating condition from
deg C (0), deg F (1).

[D21:TEMP f] Fluid temperature at
operating conditions

Set fl uid temperature at operating condition.

[D25:DENSITY UNIT] Density Unit of Flow
measurement

Select the unit of density from kg/m

3

(0), lb/cf(1), lb/

USgal(2), lb/UKgal(3).

[D26:DENSITY f] Density at normal
operation conditions

Set the density value of the fl uid at operating
condition for mass fl ow unit

[D30:OUT LIMIT] Limit value of output and
indication

Set limit value of output from 100.0% to 110.0%

[D35:BURN OUT] Indication of the output
direction at burn out

This is indication of the output direction at burn
out. Refer to Subsection 10.1.6 “Setting of Burnout
Switch” when the output direction can be changed.

[D40:SPECIAL UNIT] Change to special
fl owrate unit

No(0) : Off the function
Yes(1) : Convert the fl ow unit for USER’S UNIT

The factor to convert the fl ow unit comes
from D43. The time unit comes from C40 or
F35.

Special(2) : Convert the fl ow unit and time unit for

the USER’S UNIT.

The factor to convert the fl ow unit and
time unit comes from D43.

[D41:BASE UNIT] Indication of the base
fl owrate unit

Indication of the basic fl owrate unit when item D40
is “Yes(1)” or “Special(2)”.

[D42:USER'S UNIT] Free unit for users

Set in up to 8 alphanumeric characters when item
D40 is “Yes(1)” or “Special(2)”.
The character and sign which can be set up are the
same as C10.

[D43:CONV FACTOR] Conversion factor

Set the conversion factor when item D40 is “Yes(1)”
or “Special(2)”.
Set the conversion factor which is for both fl ow unit
and time unit in case of “Special(2)”.