4 csa – Yokogawa digitalYEWFLO (DY) User Manual

Page 160

Advertising
background image

<14. EXPLOSION PROTECTED TYPE INSTRUMENT>

14-9

IM 01F06A00-01EN

14.4 CSA

(1) Technical Data

• Explosion

Proof

Applicable Standard: C22.1-98, C22.2 No.0-M1991,

C22.2 No.0.4-04, C22.2
No.0.5-1982, C22.2 No. 25-
1966, C22.2 No. 30-M1986,
C22.2 No. 94-M1991, C22.2
No. 142-M1987, C22.2 No.
61010-1-04, ANSI/ISA-
12.27.01-2003

Certifi cate: 1166201
Type of Protection: Explosionproof for Class I, B, C

and D; Class II, Groups E, F and
G; Class III.

For Class I, Division 2 location:

“FACTORY SEALED, CONDUIT SEAL NOT

REQUIRED.”

Enclosure : Type 4X
(Integral Type Vortex Flowmeter and Remote Type
Vortex Flow Detector)

Temperature Code

Ambient

Temperature

Process

Temperature

T6 60°C

≤85°C

T5 60°C

≤100°C

T4 60°C

≤135°C

T3 60°C

≤200°C

T2 60°C

≤300°C

T1 60°C

≤450°C

Temperature Code: T6 (Remote Type Vortex Flow

Converter)

Ambient Temperature: -29 to +60°C (Integral Type

Vortex Flowmeter and
Remote Type Vortex Flow
Detector)
-40 to +60°C (Remote Type
Vortex Flow Converter)

Power Supply: 42Vdc max. (Integral Type Vortex

Flowmeter and Remote Type Vortex
Flow Converter)

Output Supply (Integral Type Vortex Flowmeter):

Current Output; 4 to 20mAdc

Pulse Output; On=2Vdc, 200mA

Off=42Vdc,

4mA

Output Signal (Remote Type Vortex Flow Detector):

Output Signal; 30Vp-p, 100

μAp-p

Input/Output signal (Remote Type Vortex Flow
Converter):

Current Output; 4 to 20mAdc

Pulse; On=2Vdc, 20mA

Off=42Vdc,

4mA

Input Signal; 30Vp-p, 100

μAp-p

Electrical Connection: ANSI 1/2 female (Special)

• Intrinsically

Safe

Type “n” and Non-incendive

Applicable Standard: C22.2 No. 0-M91, C22.2

No. 0.4-2004, C22.2 No.
157-M1987, C22.2 No. 213-
M1987, C22.2 No. 1010.1-92,
CAN/CSA-E60079-0:02,
CAN/CSA-E60079-11:02,
CAN/CSA-E60079-15:02,
ANSI/ ISA-12.27.01-2003

Certifi cate

: 1198227

Type of Protection: Ex ia IIC T4...T1 and Ex nC IIC

T4...T1

(Integral Type Vortex Flowmeter
andRemote Type Vortex Flow
Detector)

Ex ia IIC T4 and Ex nC IIC T4

(Remote Type Vortex Flow
Converter)

(Integral Type Flowmeter and Remote Type Vortex
Flow Detector)

Temperature Code

Ambient

Temperature

Process

Temperature

T4 60°C

≤135°C

T3 60°C

≤200°C

T2 60°C

≤300°C

T1 60°C

≤450°C

Ambient Temperature: -29 to +60°C (Integral Type

Vortex

Flowmeter and Remote Type
Vortex Flow Detector)

-40 to +60°C (Remote Type
Vortex Flow Converter)

Degree of Protection of Enclosure: IP67
Electrical Parameters: Ui=30Vdc, Ii=165mAdc,

Pi=0.9W

Ci=12nF,

Li=0.15mH

Electrical Connection: ANSI 1/2 NPT female
Type of Protection: Intrinsically Safe for Class I,II,III,

Div.1, Groups A,B,C,D,E,F and
G, Non- incendive for Class I,II,
Div.2, Groups A,B,C,D,E,F and
G, Class III, Div.1

Advertising