3 setting parameters using bt200, Setting parameters using bt200 -3, Important – Yokogawa digitalYEWFLO (DY) User Manual

Page 59: 1) setting flow span, Example : change fl ow span 100m, H to 150m

Advertising
background image

<7. OPERATION FOR THE BRAIN TERMINAL (BT200)>

7-3

IM 01F06A00-01EN

7.3 Setting Parameters using BT200

This section describes the setting method using a BRAIN TERMINAL (BT200). For details on the method, refer
to Section 6.3 “Parameters List” and Section 6.4 “Parameters Description”.

(1) Setting Flow Span

Example : Change fl ow span 100m

3

/h to 150m

3

/h

F0704.ai

Move the inverse video bar
to “B10 : FLOW SPAN”.

The data setup screen is
displayed.
If the security screen is
displayed, enter the security
code.

Enter “150” using the
alphanumeric keys.

Pressing the enter key,
inverse video bar is flashed.

or

The data setting can be
completed.

If data setting is missing,
press “F3” key and reset the
data.

Setting is completed.
The screen returns to the
preceding screen when the
F4” (ESC) key is pressed.

MENU

A:DISPLAY

B:EASY SETUP

C:BASIC SETUP

D:AUX. SETUP

E:METER SETUP

H:ADJUST
HOME SET ADJ ESC

PARAM

B10:FLOW SPAN

100 m3/h

B15:DAMPING

4 sec

B20:CONTACT OUT

SCALED PULSE
DATA DIAG PRNT ESC

PARAM

B10:FLOW SPAN

100 m3/h

B15:DAMPING

4 sec

B20:CONTACT OUT

SCALED PULSE
DATA DIAG PRNT ESC

SET

B10:FLOW SPAN

100 m3/h

0

DEL CLR ESC

SET

B10:FLOW SPAN

100.00 m3/h

150


DEL CLR ESC

SET

B10:FLOW SPAN

100.00 m3/h

150

print off

F2:printer on
FEED POFF NO

SET

B10:FLOW SPAN

150.00 m3/h

FEED NO OK

Enter menu item, and press
“ ” key.
Move the inverse video bar
to “B : EASY SETUP”.

IMPORTANT

After setting a parameter, keep the power on for
at least 30 seconds.
If the power of fl owmeter is turned off, a
parameter setting is released.