3. INSTALLATION

WARNING

This instrument must be installed by expert
engineer or skilled personnel. The procedures
described in this chapter are not permitted for
operators.

3.1 Installation

Precautions

(1) Ambient Temperature

Avoid an area which has wide temperature

variations. When the installation area is
subjected to heat radiation from process plant,
ensure adequate heat prevention or ventilation.

(2) Atmospheric Conditions

Avoid installing the vortex fl owmeter in a

corrosive atmosphere. When the vortex
fl owmeter must be installed in a corrosive
atmosphere, adequate ventilation must be
provided

(3) Mechanical Shock or Vibration

The

vortex

fl owmeter is of sturdy construction,

but select an area subject to minimize
mechanical vibration or impact shock. If
the fl owmeter is subject to vibrations, it is
recommended that pipeline supports to be
provided as shown in Figure 3.1.

F0301.ai

digitalYEWFLO
Vortex Flowmeter

Pipeline

Pipeline Support

Figure 3.1

Example of Pipeline Support

(4) Precautions Regarding Piping

(a) Ensure that the process connector bolts are

tightened fi rmly.

(b) Ensure that no leak exists in the process

connection pipeline.

(c) Do not apply a pressure higher than the

specifi ed maximum working pressure.

(d) Do not loosen or tighten the fl ange mounting

bolts when the assembly is pressurized.

(e) Handle the vortex fl owmeter carefully when

measuring dangerous liquids, so that the liquids
do not splash into eyes or on face. When using
dangerous gases, be careful not to inhale them.

(5) Other Considerations

• Choose a location where is suffi cient clearance

around digitalYEWFLO exist to allow such work
as routine inspections.

• Choose a location that ensures easy wiring and

piping.

3.2 Piping

Precautions

Straight Pipe Length and

Recommendations

Refer to Table 3.1 about Valve Position and
Straight Pipe Length and so on.

● Piping support

Typical vibration immunity level is 1G for normal
piping condition.Piping support should be fi xed in
case of over 1G vibration level.

● Installation direction

If a pipe is always fi lled with liquids, the pipe can be
installed vertically or at inclined angle.

● Adjacent pipes

The process pipline inner diameter should be larger
than the digitalYEWFLO inner diameter.
Use the following adjacent pipe.

Model Code

Adjacent Pipe

DY015 up to DY050

DY025/R1 up to DY080/R1

DY040/R2 up to DY100/R2

Sch40

or larger inner

diameter than

Sch40

DY080 up to DY400

DY100/R1 up to DY200/R1

DY150/R2 up to DY200/R2

Sch80

or larger inner

diameter than

Sch80