Yokogawa digitalYEWFLO (DY) User Manual

Page 44

Advertising
background image

<6. PARAMETERS>

6-8

IM 01F06A00-01EN

Item

Name

R / W

Data Range

Unit

Decimal

Point

Remark

Initial
Value

Disp. U / D

(*1)

F29 TEMP b

W -999.9 to 999.9

F28

1

Temperature of standard condition

15.0

D

L

F30 1st coef

W -32000 to 32000

1/F28

0 to 5 1st temperature coeffi cient

1.0

D

L

F31 2nd coef

W -32000 to 32000

1/F28

2

0 to 5 2nd temperature coeffi cient

1.0

D

L

F32 MASS UNIT

W kg

(0)

t (1)
Ib (2)
k Ib

(3)

Selection of mass fl ow rate unit

(0)

D

L

F35 TIME UNIT

W /s

(0)

/m (1)
/h (2)
/d (3)

Selection of time unit

1

D

L

F40 FLOW SPAN

W 0.00001 to 32000

FU+35

0 to 5 Flow span

0.5

D

L

F45 DAMPING

W 0 to 99

sec

0

Damping

4

D

L

F50 TEMP DAMPING

W 0 to 99

sec

0

Damping for temperture output

4

D

L

F52 CABLE LENGTH

W 0 to 30

m

0

Cable length for signal cable (0m in
case of integral version)

0

D

L

F55 A/OUT SELECT

W FLOW

(0)

TEMP (1)

Selection of analog output

(0)

D

L

(Indicate and Set only for F55: TEMP)

F56 TEMP 0%

W -999.9 to 999.9

D20

1

Temperture value at 0%

-40

D

L

F57 TEMP 100%

W -999.9 to 999.9

D20

1

Temperture value at 100%

250

(*2)

D

L

F58 TEMP ERR OUT

W 0%

(0)

OUT LIMIT(H)

(1)

TEMP f

(2)

Selection of themometer error output
when “F55: TEMP” is selected (A
value of OUT LIMIT(H) depend on
D30)

1

D

L

F60 SELF CHECK

R

GOOD
ERROR

Self-diagnostic message

(*1) : Available for 3.10 or greater version that can be checked in K50.
(*2) : If 7.00 or less version that can be checked in K50, Initial Value is 260.

Advertising