Yokogawa digitalYEWFLO (DY) User Manual

Page 83

Advertising
background image

<9. OPERATION VIA HART CONFIGURATION TOOL (HART 7)>

9-11

IM 01F06A00-01EN

• Simulation Setting and Correlation of Output Value

<Case A>: Without option code /MV

Simulation

Setting value

Output value

Flow rate

Total fl ow rateProcess temperatureProcess densityContact outputPV % RangeLoop Current

Flow rate

Yes

Yes

No

No

Yes

Yes

Yes

Total fl ow rate

No

Yes

No

No

No

No

No

Process

temperature

No

No

Yes

No

No

No

No

Process density

No

No

No

Yes

No

No

No

PV % Range

No

No

No

No

No

Yes

Yes

Loop Current

No

No

No

No

No

No

Yes

Yes : Simulation value or calculation result of Simulation.
No : Actual process value or parameter setting value.

<Case B>: With option code /MV
<Case B-1>: Function == “Monitor only”
<Case B-2>: Function == “Saturated Steam” or “Superheat

Steam” or “LIQUID:Mass”

<Case B-3>: Function == “Gas:STD/Normal”
<Case B-4>: Function == “Not use”

Simulation

Setting value

Output value

Flow rate

Total fl ow rate

Process temperature

Process density

Contact output

Flow rate

Yes

Yes

No

No

Yes

Total fl ow rate

No

Yes

No

No

No

Process

temperature

<Case B-1, 4>
No

<Case B-1, 4>
No

Yes

<Case B-1, 2>
Yes

<Case B-1, 4>
No

<Case B-2, 3>
Yes

<Case B-2, 3>
Yes

<Case B-3, 4>
No

<Case B-2, 3>
Yes

Process density

<Case B-1, 3, 4>
No

<Case B-1, 3, 4>
No

No

Yes

<Case B-1, 3, 4>
No

<Case B-2>
Yes

<Case B-2>
Yes

<Case B-2>
Yes

PV % Range

No

No

No

No

No

Loop Current

No

No

No

No

No

Simulation

Setting value

Output value

PV= Instaneous fl ow rate

PV= Process temperature

PV % Range

Loop Current

PV % Range

Loop Current

Flow rate

Yes

Yes

<Case B-1, 2, 3>
No

<Case B-1, 2, 3>
No

<Case B-4>
Not available

<Case B-4>
Not available

Total fl ow rate

No

No

<Case B-1, 2, 3>
No

<Case B-1, 2, 3>
No

<Case B-4>
Not available

<Case B-4>
Not available

Process

temperature

<Case B-1, 4>
No

<Case B-1, 4>
No

<Case B-1, 2, 3>
Yes

<Case B-1, 2, 3>
Yes

<Case B-2, 3>
Yes

<Case B-2, 3>
Yes

<Case B-4>
Not available

<Case B-4>
Not available

Process density

<Case B-1, 3, 4>
No

<Case B-1, 3, 4>
No

<Case B-1, 2, 3>
No

<Case B-1, 2, 3>
No

<Case B-2>
Yes

<Case B-2>
Yes

<Case B-4>
Not available

<Case B-4>
Not available

PV % Range

Yes

Yes

<Case B-1, 2, 3>
Yes

<Case B-1, 2, 3>
Yes

<Case B-4>
Not available

<Case B-4>
Not available

Loop Current

No

Yes

<Case B-1, 2, 3>
No

<Case B-1, 2, 3>
Yes

<Case B-4>
Not available

<Case B-4>
Not available

Advertising