<14. EXPLOSION PROTECTED TYPE INSTRUMENT>

14-11

IM 01F06A00-01EN

(5) Installation Diagram Intrinsically Safe

(and Note)

Intrinsically safe

+

-

+

-

+

-

+

-

+

-

+

-

+

-

+

-

+

-

+

+

-

+

[Remote type]

DYA (converter)

SUPPLY

PULSE

A

B

T (*1)

C

A

B

T

DY (flowmeter)

SUPPLY

PULSE

Safety barriers

Safety barriers

[Integral type]

DY-N (flowmeter)

Hazardous Location

Group IIC, Zone 0

Class I, II, III, Division 1,
Groups A, B, C, D, E, F and G

Non Hazardous
Location

DYC: Signal cable

(*1) Wire for T terminal
With temperature sensor type : installed
Without temperature sensor type: not installed

F1408.ai

Electrical parameters of vortex flowmeter (DY) and vortex flow
converter (DYA).
Ui (Vmax)=30 V Ii (Imax)=165 mA Pi (Pmax)=0.9W
Ci=12 nF

Li=0.15 mH

Installation requirements between flowmeter, converter and
Safety Barrier
Uo ” Ui Io ” Ii Po ” Pi Co • Ci+Ccable
Lo • Li+Lcable
Voc ” Vmax Isc • Imax Ca • Ci+Ccable
La • Li+Lcable
Uo, Io, Po, Co, Lo, Voc, Isc, Ca and La are parameters of
barrier.

WARNING

• In any safety barrier used output current

must be limited by a resistor ‘R’ such that
Io=Uo/R or Isc=Voc/R.

• The safety barrier must be CSA certifi ed.
• Input voltage of the safety barrier must be

less than 250Vrms/Vdc.

• Installation should be in accordance with

Canadian Electrical Code Part I.

• Dust-tight conduit seal must be used when

installed in class II and III environments.

• Do not alter drawing without authorization

from CSA.

Type “n” and Non-incendive

Hazardous Location

Non Hazardous
Location

Class IIC, Zone 2,
Class I, II,
Division 2,Groups A, B, C, D,
E, F and G,
Class III, Division 1

[Integral type]

[Remote type]

DY (flowmeter)

SUPPLY

PULSE

+

+

+

+

+

+

Power

Supply

Receiver

DY (converter)

DY-N (flowmeter)

SUPPLY

PULSE

+

+

+

+

+

+

Power

Supply

Receiver

A

B

T

C

A

B

T

(*1)

DYC: Signal cable

(*1) Wire for T terminal
With temperature sensor type : installed
Without temperature sensor type: not installed

Non-incendive field wire parameters of vortex flowmeter (DY)
and vortex flow converter (DYA).
Ui (Vmax)=30 Ii (Imax)=165mA Pi (Pmax)=0.9W
Ci=12nF

Li=0.15mH

Installation requirement between flowmeter, converter and
general purpose equirement.
Uo ” Ui Io ” Ii Po ” Pi Co • Ci+Ccable
Lo • Li+Lcable
Voc ” Vmax Isc ” Imax Ca • Ci+Ccable
La • Li+Lcable
Uo, Io, Po, Co Voc, Isc, Ca and La are nonincendive field
wire parameters of general purpose equipment.

F1409.ai

General Purpose

Equipment

WARNING

• The general purpose equipment must

be CSA certifi ed as the equipment which
have type n or non-incendive fi eld wire
parameters.

• Installation should be in accordance with

Canadian Electrical Code Part I.

• Dust-tight conduit seal must be used when

installed in class II and III environments.

• Do not alter drawing without authorization

from CSA.

(6) Dual Seal (Option code: /CF11, /CS11)

Dual Seal:

Certifi ed by CSA to the requirement of ANSI/ISA
12.27.01
No additional sealing required.
Primary seal failure annunciation: at the O-ring seal
portion between shedder bar and amplifi er housing.