2) burst mode setting procedure, Procedure to call up the display – Yokogawa digitalYEWFLO (DY) User Manual

Page 77

Advertising
background image

<9. OPERATION VIA HART CONFIGURATION TOOL (HART 7)>

9-5

IM 01F06A00-01EN

(2) Burst mode setting procedure

• Procedure to call up the display

DD (HART 7)
DTM (HART 7)

[Root Menu]

→ Detailed setup →

Confi gure outputs

→ HART output →

Burst Condition

→ Burst Message 1,2

or 3

→ Burst Command

F0906.ai

Burst transmition

setup

Setup of Device Valiables

Cmd9, Cmd33

Setup of Update Period,

Max Update Period

Burst

Trigger Mode ?

Setup of Burst Trigger Level

Setup of Burst Mode

Transmit to the device

Transmit to the device

Transmit to the device

Setup of Burst Command

Window
Rising
Falling

Continuous
On-change

Burst

Command ?

Cmd1, Cmd2

Cmd3, Cmd48

Setup of Burst Trigger Mode

Method: Set Burst Period

Method: Set Burst Trigger

a.

b.

c. Update Period and Max

Update Period
by "Set Burst Piriod"

d. Burst Msg Trigger Mode

by "Set Burst Trigger"

e. Burst Mode

Burst Variable Code

Burst Command

Advertising